Manufacturer Analytical
Features
Measurement range | 0-25 ppm CO |
Accuracy | < ±2 ppm under constant conditions |
Response time | T90 < 30 seconds |
Sensitivity (LDL) | 0.5 ppm CO |
Linearity | ±1% under constant conditions |
Sensor model | OSV-22-AF |
Sensor life at 25°C (77°F) and 1 atmosphere | 24 months in air 5°C (77°F) at 1 atmosphere |
Calibration interval | Certified span gas mixture of CO concentrations between 10-20 ppm with balance air. |
Inlet pressure | 800 1200 mbara (11.6-17.4 psig) |
Flow rate (constant) | 0.2-10 Nl/min (0.4-20 SCFH) |
Gas connections | Supplied with Dome adapter (A-3388) with optional BC adapter (A-3671) |
Display | 2 digit LCD 2.8 x 1.6cm (1.1 x 0.625"); resolution 1 ppm CO |
Dimensions | 6.9 x 10.4 x 3.4cm (2.72" x 4.1" x 1.35") |
Weight | 196g (7oz) |
Compensation | N/A |
Alarms | User adjustable HI and LO alarms (AII-3000 M and MHC only) |
Storage temperature | 3°C to 35°C (37.4°F to 95°F) |
Operating temperature | 0°C to 50°C (32°F to 122°F) |
Power | 2 AA alkaline batteries |
Battery Life | Up to 450 hrs of continual use |
Humidity | Non-condensing 0-95% RH |
Warm up time | 60 Seconds |